Model No.:636 Double Tilt Cooling Holder and 652 Double Tilt Heating Holder
Manufacturer:Gatan
Location:SES

The EMS also has Gatan hot and cold double tilt stages designed for the JEM-2010F and JEM-3010.

The cold stage is liquid nitrogen cooled and can reach a temperature of -170 deg C. Thermal drift after 45 minutes is less than 0.1nm/second. The maximum tilt for this stage is 18 degrees.

The heating stage can be used up to 1000 degrees C. Above 500 deg C additional water cooling is necessary. The temperature is reached within 1 minute and after 10 minutes drift rates should be less than 0.2nm/second. The maximum tilt for this stage is 8 degrees in X and 6 degrees in Y.