|
|
|
|
| Model No.: | AutoMACS |
| Manufacturer: | Miltenyi-Biotec |
| Location: | RRC-West, MSB E-6E |
|
Occasionally, researchers need very large numbers of purified cells which
means that sorting for purity with flow cytometry is impractical because
it would take too long. To accommodate these people, the RRC purchased a
Miltenyi-Biotec automatic magnetic bead cell separator (AutoMACS), which
can sort millions of cells in minutes. This instrument was installed in
December of 1999 and has been used often as a presorter for very rare
cell populations. In this instance, the few desired cells were first
separated from the millions of unwanted cells with the AutoMACS, and then
sorted by flow cytometry in order achieve high purity. The AutoMACS is
also very useful in instances where time is of essence and extremely high
purity is not an issue. This instrument makes the cell purification with
magnetic beads simple and fast. Manual magnetic bead separations take up
to two or four times longer than using this instrument and pressure / flow
rates on manual separation columns are hard to control. The drawbacks for
this instrument are 1. it can only be used for markers that are on the cell
surface (intracellular markers such a GFP would have be purified with flow
cytometry) and 2. the purity is not as good (75%-95% purity) when compared
with flow cytometry sorting (97%-99% purity). Purity is compromised when
using the AutoMACS positive selection mode because your desired cells will
be contaminated with dead cells. The depletion mode on the AutoMACS will
generate cell preparations of greater purity but you would need to have
beads that will react with all of your unwanted cells. For more
information on the AutoMACS, contact the University of Chicago web page at:
http://iacf.bsd.uchicago.edu/FlowHome/InstrumentPics/Automacs.htm
At present there is no instrument charge for the use of this instrument. Only the cost of materials (columns and buffers) is charged to the investigator.
|